Characteristic
Rated output AC: 0.050 KV-5.000 KV
Rated output DC: 0.050 KV-6.000 kV
Touch key control
Large screen disy
English and Chinese operation interface
Standard RS232. Handler interface
Specifications
| Test function | Ac voltage resistance, DC voltage resistance, insulation resistance |
| Interface | RS232/Handler |
| Arc detection | Level 9 (AC/DC) |
| Other functions | List Settings
Discharge function 5 "True color LCD touch screen |
| Ac withstand voltage | |
| Rated output | AC 5.000kV/10mA |
| Output voltage | 0.050kV-5.000KV AC
Resolution: 1V |
| Voltage accuracy | 2% of reading + 5 digits |
| Voltage adjustment rate | 2% of reading +5V
(from empty to full) |
| Output frequency | 50Hz or 60Hz -- is optional |
| Measure the current | 0.001mA-10.00mA
Resolution: 0.001/0.01 |
| Current accuracy | 2percent x reading 5 words |
| Voltage rise time | 0.1s-999.9s |
| Test time | 0.2s-999.9s |
| Voltage drop time | 0.1s-999.9s |
| Dc voltage resistance | |
| Rated output | DC 6kV/5mA |
| Output voltage | 0.050kV-6.000kV DC
Resolution: 1V |
| Voltage accuracy | 2% of reading + 5 digits |
| Voltage adjustment rate | 2% of reading +5V |
| Current measurement | 0.1μA-5.000mA |
| Current accuracy | 2% of reading +5 digits |
| Voltage rise time | 0.1s-999.9s |
| Test time | 0.2s-999.9s |
| Voltage drop time | 0.1s-999.9s |
| Insulation resistance | |
| Rated output | DC 1kV/10GΩ |
| Output voltage | 0.050kV-1.000kV DC
Resolution: 1V |
| Voltage accuracy | 1% of reading + 2 digits |
| Measuring range | 0.1MΩ-10GΩ |
| Resistance accuracy | <500V:
0.1MΩ-1GΩ, Accuracy:±(10% of reading +5 digits ) 1GΩ-10GΩ, for reference only ≥ 500V: 1MΩ-1GΩ, Accuracy:±(5% of reading +5 digits) 1GΩ-10GΩ, Accuracy:±(10% of reading +5 digits) |
| Test time | 0.2s-999.9s |
| Power requirements | 198V ~ 242V 50/60Hz
400VA |
| Size and weight | 270mm(W)x130mm(H)x450mm (D)
10kg |