Product specifications: Fluke 6500-2 electrical safety tester
| Technical index | ||
| 测试Technical index | The accuracy indicator of the disy range is defined as ±(percent of the number of words read), applicable to 23 °C ±5 °C,≤75percent RH. | |
| At 0 °C to 18 °C and 28 °C to 40 °C, the accuracy index decreases per °C (0.1 x accuracy index). | ||
| The range meets the service operation error requirements specified in EN61557-1: 1997,EN61557-2: 1997,EN61557-4: 1997,EN61557-6: 1997,DIN VDE0404-2. | ||
| Power-on test | Test indicates reverse L-N, lack of PE, and measures supply voltage and frequency | |
| Disy range | 90v - 264v | |
| Accuracy @ 50 Hz | ±(2percent 3 words) | |
| Resolution | 0.1 V | |
| Input impedance | > 1 MΩ // 2.2 nF | |
| Maximum input voltage: | 264 V | |
| Ground connection test (Rpe) | Disy range | 0 to 19.99 Ω |
| Accuracy (after grounding test zero) | ±(5percent 4 words) | |
| Resolution | 0.01 Ω | |
| Test the current | 200 mA AC -0% +40%,1.99 Ω
10 A AC ±20%,25 mΩ @ 230 V |
|
| Open-circuit voltage | > 4 V AC、<24 V AC | |
| Ground test zero setting | You can subtract up to 1.99 Ω | |
| Insulation test (Riso) | Disy range | 0 to 299 MΩ |
| Accuracy | ±(5percent 2 words),0.1 to 300 MΩ | |
| Resolution | 0.01 mΩ (0 - 19.99 mΩ)——
0.1 mΩ (20 - 199.9 mΩ)—— 1 mΩ (200 - 299 mΩ)—— |
|
| Test voltage | 500 V DC-0percent 25percent @ 500 kΩ load, or 250 V dc-0percent 25percent @ 250 kΩ load | |
| Test the current | 比;1毫安@ 500 kΩ,<15毫安@ 0 Ω—— | |
| Automatic discharge time | <0.5 s,1 μF | |
| Maximum capacitive load | Operation up to 1μF | |
| Contact current test | Disy range | 0 to 1.99 mA,AC |
| Accuracy | ±(4percent 2 words) | |
| Resolution | 0.01 mA | |
| Internal resistance (through the probe) | 2 kΩ | |
| Measuring method | Probe * | |
| Instead of leakage current testing | Disy range | 0 to 19.99 mA,AC |
| Accuracy | ±(2.5percent 3 words) | |
| Resolution | 0.01 mA | |
| Test voltage | 100 V AC ±20% | |
| Load/leak test: Load current | Disy range | 0 A to 13 A |
| Accuracy | ±(4percent 2 words) | |
| Resolution | 0.1 A* | |
| Load/leakage current test: load power | Disy range230 VPower supply | 0.0 VA- 3.2 kVA |
| Accuracy | ±(5percent 3 words) | |
| Resolution | 1va (0 ~ 99va),0.1 kVA (>1.0 kVA)*—— | |
| Load/leakage current test: Leakage current | Disy range | 0 to 19.99 mA |
| Accuracy | ±(4percent 4 words) | |
| Resolution | 0.01 mA* | |
| PELV测试 | Accuracy @ 50 Hz | ±(2percent 3 words) |
| Overload protection | 300 V rms | |
| Alarm threshold | 25 V rms | |
| RCD test: Trip current | Operating error | ±10% |
| Nominal value | 30 mA | |
| Accuracy | ±5% | |
| RCD test: Trip time | Standard requirement | 61557第6章;额定Test the current允差为0%至+10% |
| Operating error | ±10% | |
| RCD type | Ac Universal 30 mA | |
| Disy range | 310 ms | |
| Resolution | 0.1 ms | |
| Accuracy | 3 ms | |
| Trip time limit @100percent (30 mA) | 300 ms | |
| Trip time Limit @500percent (150 mA) | 40 ms | |
| * The electrical device under test is powered by the mains power supply | ||
| Environmental indicators | |
| Operating temperature | 0 to 40 °C |
| Relative humidity | non-condensing 10 °C |
| 95%,10 - 30°C | |
| 75%,30 to 40 °C | |
| Safety index | |
| Safety level | In accordance with the requirements of EN61010-1 third edition |
| CAT II,300 V, contamination level 2 | |
| 机械及通用Technical index | ||
| Dimensions (length x width x height) | 200 x 275 x 114 mm | |
| The weight of | 3.13 kg | |
| Power supply | 230v 10% - 15%, 50hz±2hz或(2.065 - 2:10 0v 10% - 15%, 50hz±2hz)—— | |
| Power consumption (tester) | 13 W Typical (standby)
60 W Max, 25A ground test period |
|
| Store | Temperature | -10 - 60°c |
| Corrosion | 70 °C @ 95percent Rh, up to 5 days | |
| Working altitude | 0 to a maximum of 2000 m | |
| Seal | IP-40 (shell),IP-20 (connector) | |
| EMC | In line with EN61326-1 standard, portable | |
| EMI immunity | 3 V/m | |