Product specification: Fluke 51-II single channel contact thermometer
Technical index | |
Temperature accuracy | Above -100 °C:J,K, T,E type: ±[0.05percent 0.3°C]*R and S type: ±[0.05percent 0.4°C]* Below -100 °C:J,K,E type: ±[0.20percent 0.3°C]*T type: ±[0.50percent 0.3°C] |
Temperature range | J-type: -210 °C to 1200 °CK type: -200 °C to 1372 °CT type: -250 °C to 400 °CE type: -150 °C to 1000 °C0 °C to 1767 °C * |
Temperature scale | ITS-90 |
Applicable standards | NIST-175 |
Disy resolution | 0.1 °C, 0.1 K <10001°C, 1 K >=1000 |
Note | * Only Fluke 53 II B and 54 II B thermometers can measure N,R and S thermocouples |
Environment parameters: | |
Operating temperature | -10°c - 50°c |
Storage temperature | -40°c - 60°c |
Humidity (no condensation) | 0°C- 35°C- 70%;0°C- 50°C |
Safety technical parameters | |
Class of overvoltage | CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2 |
Institutional certification | CE, CSA, TUV (to be approved) |
Mechanical and general parameters | |
Size | 173 mm L x 86 mm W x 38 mm |
The weight of | 400 g |
Battery | 3 节 AA 型号Battery,典型的Battery寿命为1000小时 |