Product specifications: Fluke 719 portable automatic pressure calibrator
| General technical specification | ||
| Storage temperature | -30°c - 60°c | |
| Operating temperature | -10°c - 55°c | |
| Relative humidity | 95%(-10°C - 30°C) | |
| 75%(30°C - 40°C) | ||
| 45%(40°C - 50°C) | ||
| 35%(50°C - 55°C) | ||
| Impact | 1 meter drop test | |
| Safety level | CAN/CSA C22.2 010.1:1992 | |
| Electromagnetic compatibility (EMC) | EN50082-1:1992 and EN55022:1994 Class B | |
| Dimensions (height x width x length) | 60 x 87 x 210 mm | |
| (2.19 x 3.41 x 8.28 in) | ||
| The weight of | 912 g (2 lb) | |
| Battery | 两节 9 V 碱性Battery | |
| Battery寿命 | 12 hours (12 mA inflow at 500 Ω load), 200 pump cycles at 100 psi, 500 pump cycles at 30 psi | |
| Warranty period | Pressure calibrator three years, air pump one year | |
| Technical specifications | ||
| Fluke-719 30G | The range of | -12 - 36psi,- 850mbar - 2.4 bar |
| Resolution | 0.001 psi,0.1 mbar | |
| Accuracy | The range of的 0.025 %,6 个月
The range of的 0.035 %,1年 |
|
| Fluke-719 100G | The range of | -12 - 120psi,- 850mbar |
| Resolution | 0.01 psi,1 mbar | |
| Accuracy | The range of的 0.025 %,6 个月
The range of的 0.035 %,1年 |
|
| mA | The range of | 0 to 24 mA |
| Resolution | 0.001 mA | |
| Accuracy | 2 counts of 0.015percent of the reading | |
| Maximum load 1000 Ω, 750 Ω ohm-- in HART mode | ||
| Loop power supply | The range of | Dc 24 V |
| Resolution | N/A | |
| Accuracy | ±10% | |
| Pump parameters | ||
| Maximum formation pressure | 120psi (8bar)(100psi | |
| Pumping | -30psi less than 15 seconds | |
| -100psi less than 45 seconds | ||
| 1. With 1 meter 1/8 in test hose | ||
| The number of pump cycles between repairs | 5000 | |
| Total pump life | 20,000 calibration cycles | |